Test purpose
Verify the detection capability of the IQM System for a variety of clinically relevant delivery errors.
Test method
3 clinical IMRT plans were modified: Brain, Prostate, Head & Neck. All errors had a clinical effect: DVH parameters of either the targets or organs at risk changed by a few percentages.
Conclusion
IQM showed a very high error detection rate. Using a watch level of 2% and an action level of 3% IQM would have detected all but one of the induced errors.
This test report is based on a publication created by University Clinic, Würzburg (Germany).