Test method
The IQM error sensitivity was evaluated by introducing deviations in 4 clinical H&N VMAT plans:
3, 5 and 10 % errors on total delivered MUs
and 3, 5 and 10 mm MLCs shift introduced with a Matlab script.
The cumulative IQM checksum value was measured and the percentage difference was calculated with respect to the non-modified plan.
Test results
Both the MLC shift errors as well as the MU variations were detected by IQM.
IQM shows a linear response with dose (R2=0.9995).
Conclusion
IQM shows a good ability in detecting real-time errors and seems to be a good candidate to save time during the plan QA process.
This test report is based on a publication created by Azienda Provinciale per i Servizi Sanitari, Trento (Italy).
Please click on the „Download“ button to download the complete publication.