Test purpose
Test the inter-fractional (long-term) repeatability of the IQM signal.
Test method
Inter-fraction repeatability was checked by delivering three IMRT treatments for more than seventy times in a period of thirty days. The chosen treatments were: a head & neck (H&N), a prostate and a test IMRT treatment. For each segment of the IMRT treatment sigma and M were evaluated over the number of repetitions (21 for H&N; 30 for prostate and „Test“ IMRT). In H&N and prostate tests gantry rotation and small fields of different shape and size were considered.
Conclusion
IQM provides a high long-term signal reproducibility for complex IMRT plans.
This test report is based on a publication created by University of Florence, Florence (Italy).