Test purpose
Evaluate the correlation between changes in IQM signal, Gamma-Index and Dose Volume Histogram.
Test method
The sensitivity of detecting small delivery errors was checked by inducing seven types of errors in the IMRT clinical plans for head & neck, prostate and the IQM test plan by modifying the number of delivered MU (between 1 and 3 per beam) and by introducing deviations in linac leaf positions mimicking an MLC bank error as closing and opening one or both banks. The influence of those errors on the IQM signal, the change in Gamma Index as well as the influence on the dose volume histogram (PTV V95%) was analyzed.
Conclusion
A good correlation between IQM signal variations and 3D Gamma, 2D Gamma and DVH parameters has been observed.
This test report is based on a publication created by University of Florence, Florence (Italy).